NDIR / TOC is detected directly over Carbon Dioxide and not via a 3rd parameter (conductivity). Therefore no limit to conductivity.
Available for two ranges of TOC
9 months UV lamp life guarantee / 50% longer life
2 sample inlets / 2 points of the loop can be measured, so double efficiency
Remote service / online service diagnosis
UV oxidation (full) & NDIR
20 mL (low range), 5 mL loop (high range)
2 to 5000 ppb (low range), 15 ppb to 20 ppm (high range)
USP <643>, EP 2.2.44, DIN EN 1484
DIN 32645, 21 CFR part 11, IP 45
The scheme shows the principle of uniTOC with to different pathways for TIC and TOC and the principle of measuring the TOC value.
If the feed pressure is higher than 1.5 bar a pressure transducer is required.
Validation documents IQ/OQ for uniTOC.
To use compressed air instead of oxygen or nitrogen.
Keeps the temperature below 50 °C degrees.
Do you have any questions?
We would be more than happy to help you.
Technical Specifications |
|
---|---|
method & detection | UV oxidation & NDIR (enhanced oxidation; configurable with reagent) |
sample volume | 20 mL loop (low range), 5 mL loop (high range) |
period of measurement | 6 to 15 min |
measurement range | 2 to 5000 ppb (low range), 15 ppb to 20 ppm (high range) |
detection accuracy | ~3 ppb or 2 % (whatever is greater) |
ports | 2 for samples | 1 for reagent |
related norms | USP <643>, DIN 32645, EP 2.2.44, DIN EN 1484, CFR 21 part 11, IP 45 |
ambient parameters | relative humidity: 10 to 95 %, temperature: 10 to 45 °C |
interface ports | analog (4 – 20 mA) |
dimensions, weight, power | 400 x 300 x 600 mm, 24 kg, 110 – 230 V, 50 Hz, 70 to 105 Watt |
Medium Requirements | |
---|---|
type of sample medium | from ground water to ultra pure water |
sample temperature | 10 to 50 °C |
inlet pressure | < 1.5 bar |
compressed gas | air, oxygen, nitrogen | free of carbon dioxide |
inlet pressure of gas | nitrogen: 1 to 1.5 bar | oxygen: 3.5 to 5 bar |
Contact Information
LOCATIONS
Office 3, Dynamic Business Center, Abdulkambar, Hor Al Anz, DUBAI, UAE
Get Directions To Our Office
P.O.Box: 55879
info@fmit-instruments.com
Tel: (+971) 4 5284003
© 2023 FALCON ELECTRONIC PRECISION NSTRUMENTS TRADING CO. L.L.C FMIT