Low sample volume / lower cost and waste
9 months UV lamp life guarantee / 50% longer life
Continuous measurement (every 2 seconds) / fast reaction
Offline measurement / grab samples can be measured
Up to 3 sample inlets / very efficient
Remote service / online service diagnosis
UV-Oxidation & conductivity
ca. 14 mL/min
0.5 to 1000 ppb
USP <643>, EP 2.2.44, DIN EN 1484
The scheme shows the principle of measuring the TOC via conductivity with miniTOC.
The miniTOC can be ordered as configuration with 4 inputs (miniTOC pro). With an additional software upgrade option the user can measure up to 3 loops sequentially.
An offline configuration is available to analyze single samples.
By using the offline mode of the miniTOC pro the system is also upgradable with an autosampler with 15 positions for 100 mL vials.
If the medium temperature is higher than 50 °C a heat exchanger is required.
If the medium pressure is higher than 1 bar a pressure transducer is required.
Software with 21 CFR part11 conformity for pharma applications.
The housing protects the miniTOC from dust deposits and water jets
Initiates alarm in case of inlet flow is zero.
Validation documents IQ/OQ and PQ for miniTOC (conform and non conform with 21CFR part 11 software.
extended conductivity of sample up to 10 µS/cm.
Do you have any questions?
We would be more than happy to help you.
Technical Specifications |
|
---|---|
method & detection | UV oxidation (partial) & conductivity |
sample volume flow | approx. 14 mL/min |
time of measurement | 2 sec (online mode) | 1 min (offline mode) |
measurement range | 0.5 to 1000 ppb |
detection accuracy | ~0.5 ppb or 1 % (whatever is greater) |
ports | 1 for sample, 1 for SST (basic) | 1 for sample, 3 for SST (pro) |
related norms | USP <643>, EP 2.2.44, DIN EN 1484 |
ambient parameters | relative humidity: 10 to 95 %, temperature: 10 to 45 °C |
interface ports | analog, according to Namur NE43 |
dimensions, weight, power | 300 x 200 x 500 mm, 14 kg, 110 – 230 V, 50 Hz, 60 Watt |
Medium Requirements | |
---|---|
type of sample medium | pure or ultra pure water |
medium conductivity | < 2 µS/cm (upgradable to < 10 µS/cm) |
inlet pressure | < 1 bar |
sample temperature | < 50 °C |